NEWS
Treatment method when microorganisms in the good price and quality WFI Distribution Loop exceed the correction limit
- Categories:News
- Author:
- Origin:
- Time of issue:2022-07-12
- Views:0
(Summary description)After the hydrogen peroxide solution has been drained, restart the good price and quality WFI Distribution Loop to produce water for injection into the tank, continue to repeat the cycle and discharge, repeat the cycle and discharge twice and take samples at the sampling port of the good price and quality WFI Distribution Loop system. At this time, the PH value is displayed. The properties and conductivity are within the specified range. After the disinfection is completed, the storage tank, general delivery, general circuit and key use points of the good price and quality WFI Distribution Loop can also be sampled for complete testing.
Treatment method when microorganisms in the good price and quality WFI Distribution Loop exceed the correction limit
(Summary description)After the hydrogen peroxide solution has been drained, restart the good price and quality WFI Distribution Loop to produce water for injection into the tank, continue to repeat the cycle and discharge, repeat the cycle and discharge twice and take samples at the sampling port of the good price and quality WFI Distribution Loop system. At this time, the PH value is displayed. The properties and conductivity are within the specified range. After the disinfection is completed, the storage tank, general delivery, general circuit and key use points of the good price and quality WFI Distribution Loop can also be sampled for complete testing.
- Categories:News
- Author:
- Origin:
- Time of issue:2022-07-12
- Views:0
The good price and quality WFI Distribution Loop may generate some microorganisms after being used for a period of time. In this case, a 0.5 percents hydrogen peroxide solution can be prepared to clean the WFI Distribution Loop system before sterilization. Preparation method: Hydrogen peroxide is analytically pure, and water for injection is added to prepare a 0.5% (volume concentration) hydrogen peroxide solution.
The TOC detector should be turned off when the good price and quality WFI Distribution Loop is being sterilized with hydrogen peroxide. Then turn on the pipeline pump to be cleaned, stop the pump after circulating for about an hour, and then open the sewage outlet at the bottom of the WFI Distribution Loop tank to drain the hydrogen peroxide solution.
After the hydrogen peroxide solution has been drained, restart the good price and quality WFI Distribution Loop to produce water for injection into the tank, continue to repeat the cycle and discharge, repeat the cycle and discharge twice and take samples at the sampling port of the good price and quality WFI Distribution Loop system. At this time, the PH value is displayed. The properties and conductivity are within the specified range. After the disinfection is completed, the storage tank, general delivery, general circuit and key use points of the good price and quality WFI Distribution Loop can also be sampled for complete testing.
Scan the QR code to read on your phone
Relevant information
Rigorous testing
ust the quality of products
24/7 Online services
High quality after-sales service
On time delivery
Strong production capacity, sufficient stock
Quality qualification
Good quality price is lower
CONTACT US
Shandong Eagle Machinery Co.,Ltd
Address: Block B, Headquarters Base ABP, No.2600,
Zhuangjian Road.Weifang, Shandong, P.R.C.
Tel: +86-536-8800237
Email: Hrs@sdwemac.com
WhatsApp:
Naomi: 0086-15315252862
Crystal Zhao: 0086-15053106839
Copyright © Shandong Eagle Machinery Co.,Ltd. Powered by www.300.cn 鲁ICP备19023047号-4